Problems with the detector? Sometimes the problem isn't in the sample.

This article addresses the most common HPLC detector failures and how to identify their source. It distinguishes between electronic problems, which should be resolved with the manufacturer, and mechanical or optical failures, which typically focus on the flow cell. Warning signs such as bubbles, contamination, or high pressure are also detailed, and critical maintenance aspects are mentioned, such as lamp status or parameter settings such as gain and response time. The text provides practical advice for diagnosing and preventing errors that directly affect the quality of chromatographic analysis.


Cuando algo falla en el detector, el síntoma puede ser sutil o llamativo: una línea base ruidosa, picos inestables o una caída inesperada en la sensibilidad. Lo primero es saber de dónde viene el problema, ya que suele tener dos orígenes posibles: electrónico o mecánico/óptico.

Los fallos electrónicos —como errores de conexión o circuitos dañados— deben ser atendidos directamente por el fabricante del equipo. Pero muchos de los problemas más frecuentes están relacionados con la celda de flujo, el corazón óptico del detector.

Señales de alerta que pueden indicar un problema en la celda:

  • Burbujas de aire atrapadas

  • Contaminación interna

  • Fugas o presión excesiva

  • Daños por mal manejo o desgaste

Todo esto puede traducirse en una respuesta reducida del detector, fluctuaciones en la línea base o incluso picos espurios. Además, algunos detectores, como los de índice de refracción, utilizan celdas especialmente sensibles a la presión: si se exceden los valores recomendados por el fabricante, el cristal de la celda puede romperse fácilmente.

Tampoco hay que olvidarse del estado de los componentes internos del detector: una lámpara envejecida, un ajuste incorrecto del tiempo de respuesta o una ganancia mal configurada también pueden comprometer la calidad del análisis.

 

 


Artículos recomendados

See all
When a gram is too much: real precision in the food industry
When a gram is too much: real precision in the food industry
Discover why in the food industry a small error can affect the entire process and how OHAUS Explorer scales allow you to work with real precision and without uncertainty.
Comparison of OHAUS moisture analyzers in the food industry: MB32 vs MB62 vs MB92 vs MB120
Comparison of OHAUS moisture analyzers in the food industry: MB32 vs MB62 vs MB92 vs MB120
Comparison of OHAUS moisture analyzers focused on the food industry. Discover which model to choose to improve quality, control processes, and reduce losses.
Eppendorf Research® Neo 3: precision, ergonomics and total control in professional pipetting
Eppendorf Research® Neo 3: precision, ergonomics and total control in professional pipetting
In environments where a deviation of microliters can compromise the validity of an entire test, the pipette ceases to be a simple instrument and becomes a critical element of the quality system. The Eppendorf Research® Neo 3 is positioned precisely at this critical point, where nominal accuracy is no longer sufficient and what truly matters is reproducibility under real-world working conditions.